Tecal 425 Technical
data
|
|
Minimum temperature above
ambient
|
20 °C
Above ambient
|
Cooling rate 400 °C
to 100°C
|
25 min
|
|
Maximum temperature
|
425 °C
|
Programmable ramp rate °C/min
|
0.1 to 25 on H version
|
|
Temperature accuracy in
measuring zone
|
+/0.3 °C
|
¼" reference hole in
block
|
Yes
|
|
Temperature uniformity in
measuring zone
|
+/-0.2 °C1
|
Switch test
|
On H version
|
|
Measuring zone
|
0 to 50mm from base of well
|
Fan cooling
|
Automatic
|
|
Temperature stability after 10
mins
|
+/-0.05 °C
|
Comms port, 9 way D-type
|
Full bi-directional RS232
|
|
Display resolution
|
0.1 °C
or 0.1°F
|
Weight (lbs)
|
20.0
|
|
Heating rate, 20 °C
to 400°C
|
15 min
|
Dimensions
H x W x D mm
|
285 x 190 x 426
|
|
1
At 400°C
|