Tecal 100
RTD Calibrator
Techne Home
|
High
Accuracy |
|
Large
Display for easy reading |
|
Programmable
Offset, Step and Ramps |
|
Rechargeable
Internal Battery Pack |
|
The Tecal
100 is a hand-held instrument that measures and simulates RTD's with an
accuracy of ±0.05°C/0.09°F.
Compact, easy to use and with a long life
rechargeable battery pack, it is the solution to the problem of high
accuracy RTD measurement and simulation in the plant and lab.
Calibration features include programmable offset,
steps and ramps.
Use as a high accuracy thermometer by entering
the ice-point or Ro value of your RTD.
|
|
Tecal
100 Features |
|
Accuracy
±0.05°C/0.09°F |
|
Resolution
0.01°/Ω |
|
Ohms
accuracy ±0.03Ω |
|
Read
and Simulate 2, 3 and 4 wire systems |
|
Display
readings in °C,°F, Ω & K |
|
Ramp
and Step functions |
|
Offset
for ice-point RO values and use as an accurate thermometer |
|
Select
from the following probe types: PT100 (3.85 & 3.916
alpha), PT200 (3.85 alpha), PT500 (3.85 & 3.916 alpha), PT1000 (3.85 alpha), Ni60 (6.72
alpha), and Ni120 (6.72 alpha |
Accuracy Specifications
Standard RTD types ( Non-standard RTD types are user
programmable)
| |
|
Celsius |
Fahrenheit |
Element |
Alpha
Coeff. |
Range |
Error |
Range |
Error |
Pt
100 DIN |
.00385 |
-200
to 250 |
0.05 |
-330
to 480 |
.010 |
|
Pt
100 DIN |
.00385 |
250
to 850 |
0.07 |
480
to 1560 |
.013 |
|
Pt
100 US |
.00385 |
-100
to 250 |
0.05 |
-150
to 480 |
.010 |
|
Pt
100 US |
.00385 |
250
to 457 |
0.07 |
480
to 850 |
.013 |
|
Pt
200 DIN |
.00385 |
-200
to 300 |
0.05 |
-330
to 570 |
.010 |
|
Ni
120 |
.00385 |
-100
to 200 |
0.05 |
-150
to 390 |
.010 |
|
Ni
1000 |
.00385 |
-100
to 200 |
0.05 |
-150
to 390 |
.010 |
|
Resistance |
|
Range
Ω |
Monitor
Error |
Generator
Error |
|
20
to 400 |
0.03Ω |
0.03Ω |
|
400
to 800 |
0.10Ω |
0.10Ω |
|
800
to 1500 |
0.20Ω |
0.20Ω |
|
1500
to 2600 |
0.50Ω |
0.50Ω |
General Information
|
Total non
linearity for °C, °F and K as Alpha factor |
Operating
temperature range 10 to 50°C |
|
Accuracy -
Error +/-1 digit |
Battery
life> 30 hours (Ni-cad rechargeable battery) |
|
Display
updates every 0.6 seconds |
Normal mode
50/60hz rejection 100db |
|
Case: Impact
resistant ABS |
Size:
6"h x 3.5"w x 1.75 d |
Ordering Information
Catalog # |
Description |
3024000 |
Tecal
100, 120 volt |
3024050 |
Tecal
100, 240 volt |
|