Tecal 100 RTD Calibrator                                                        Techne Home

High Accuracy
Large Display for easy reading
Programmable Offset, Step and Ramps
Rechargeable Internal Battery Pack

     
The Tecal 100 is a hand-held instrument that measures and simulates RTD's with an accuracy of ±0.05°C/0.09°F.

Compact, easy to use and with a long life rechargeable battery pack, it is the solution to the problem of high accuracy RTD measurement and simulation in the plant and lab.

Calibration features include programmable offset, steps and ramps.

Use as a high accuracy thermometer by entering the ice-point or Ro value of your RTD. 

 

Tecal 100 Features

Accuracy ±0.05°C/0.09°F
Resolution 0.01°/Ω
Ohms accuracy ±0.03Ω
Read and Simulate 2, 3 and 4 wire systems
Display readings in °C,°F, Ω & K
Ramp and Step functions
Offset for ice-point RO values and use as an accurate thermometer
Select from the following probe types: PT100 (3.85 & 3.916 alpha), PT200 (3.85 alpha), PT500 (3.85 & 3.916 alpha), PT1000 (3.85 alpha), Ni60 (6.72 alpha), and Ni120 (6.72 alpha

Accuracy Specifications
Standard RTD types ( Non-standard RTD types are user programmable)

    Celsius Fahrenheit

Element

Alpha Coeff.

Range

Error Range Error

Pt 100 DIN

.00385 -200 to 250 0.05 -330 to 480 .010
Pt 100 DIN .00385 250 to 850 0.07 480 to 1560 .013
Pt 100 US .00385 -100 to 250 0.05 -150 to 480 .010
Pt 100 US .00385 250 to 457 0.07 480 to 850 .013
Pt 200 DIN .00385 -200 to 300 0.05 -330 to 570  .010
Ni 120 .00385 -100 to 200 0.05 -150 to 390 .010
Ni 1000 .00385 -100 to 200 0.05 -150 to 390 .010

Resistance

Range Ω Monitor Error Generator Error
20 to 400 0.03Ω 0.03Ω
400 to 800 0.10Ω 0.10Ω
800 to 1500 0.20Ω 0.20Ω
1500 to 2600 0.50Ω 0.50Ω

General Information

Total non linearity for °C, °F and K as Alpha factor Operating temperature range 10 to 50°C
Accuracy - Error +/-1 digit Battery life> 30 hours (Ni-cad rechargeable battery)
Display updates every 0.6 seconds Normal mode 50/60hz rejection 100db
Case: Impact resistant ABS Size: 6"h x 3.5"w x 1.75 d

Ordering Information

Catalog #

Description

3024000

Tecal 100, 120 volt

3024050

Tecal 100, 240 volt

 

To Place an Order, or to Receive any Additional Information, Please e-mail:  process@bcgroupintl.com, call (314) 638-3800 or fax (314) 638-3200