Techne PRT's
Techne Home
Our PRT's offers proven and reliable performance for your
critical temperature measurements. They can be used with our Tecal Accu-Temp indicator or any comparable model. Both models include an ITS-90 calibration
(NIST traceable) certificate with coefficients and resistance vs. temperature
table. The WSP660 gives an SPRT level of performance and specifications at an affordable price, while the 300 deg.C model is an Industrial RTD designed for lower accuracy measurements.
|
|
660° C Working Standard PRT |
 |
Model WSP660 PRT's are assembled using our special wire wound RTD
elements, assembled into inconel sheaths. The WSP660 PRT has similar
accuracy and long-term stability specifications you would find in much
higher priced Primary standard PRT's. The WSP660 PRT can be
used as a reference standard in calibration baths, furnaces, fluidized baths, dry block
calibrators and other any other temperature calibration work requiring a
high level of accuracy and repeatability. Combine with |
|
Specifications
Resistance at 0° C: 100 ohms
Temperature range: -200 to 660°C
Calibrated Temperature range: -30 to 660°C
Temperature coefficient:
0.003925
Ω
/Ω /° C
minimum
Stability: R0 typical drift < 0.03°C after 500 hours at 660° C
Thermal shock: R0 typical drift < 0.010° C
after 100 times
from 25 to 660°C
Dimensions: 0.25" dia. x 15" L
inconel sheath as standard,
custom sizes available
Tolerance at 0° C: ± 0.06%, IEC DIN class A
Lead Wires: 4 wire Teflon insulated, stranded 22 AWG copper
x 6'
L
|
Calibration
Each WSP660 includes an ITS-90 NIST traceable
calibration with a+, b+, c+, a- and b- coefficients and R vs. T table.
Calibrated uncertainty (K=2)
-20°C ±0.052°C
0°C ±0.036°C
200°C ±0.042°C
420°C ±0.050°C
660°C ±0.070°C
(Calibrations down to -196°C
available upon request.)
|
|
300° C Working Standard PRT |
 |
The model TI913 RTD is
a good choice for lower temperature range work at a lower accuracy. The
sheath is of stainless steel construction with a mineral insulated .385
alpha RTD. The TI913 can be used as a working standard in calibration
baths, furnaces, fluidized baths, dry block calibrators or for any other
temperature calibration work. |
|
Specifications
Resistance at 0° C:
100 ohms
Temperature range: -200 to 300°C
Calibrated Temperature range: -30 to 300°C
Temperature coefficient:
0.00385
Ω
/Ω /° C
minimum
Stability:
Meets IEC stability specifications
after 250 hours exposure to extremes of temperature range. Typical drift
is less than 0.05°C (0.02Ω)
at 0°C.
Repeatability:
Meets
IEC requirements. Typical shift less than 0.05°C (0.02Ω)
at 0°C after ten cycles over range.
Dimensions: 0.25" dia. x 12" L
inconel sheath as standard,
custom sizes available
Tolerance at 0° C: ± 0.06%, IEC DIN class A
Lead Wires: 4 wire Teflon insulated, stranded 22 AWG copper
x 6'
L
|
Calibration
Each TI913 RTD includes an ITS-90 NIST traceable
calibration with a+, b+, a- and b- coefficients and R vs. T table.
Calibrated uncertainty (K=2)
-20°C ±0.076°C
0°C ±0.068°C
100°C ±0.071°C
300°C ±0.072°C
(Calibrations down to -196°C
available upon request.)
|
|