TC-100                                                                                       Martel Home
Thermocouple Calibrator
COMMON FEATURES
High accuracy - ±0.3°C (Type J T/C - all errors combined)
Ten (10) common T/C types plus mV
Accepts both T/C mini-plug and bare T/C wires
Simple decade control of output
Three (3) setpoints for each T/C type
MIN/MAX recall in measure mode
Input protection to 240 VAC
The Martel Electronics TC-100 Thermocouple Calibrator provides high accuracy source and measurement of ten common thermocouples, as well as mV. The TC-100's accuracy of ±0.3°C for Type J T/Cs includes all errors, at resolutions of ±0.01°C or °F in measure mode and ±0.1°C or °F in source mode. Features including MIN/MAX recall in measure mode, three setpoints per thermocouple range, a large knob for decade control of the output in source mode, and the ability to accept bare T/C wires in addition to mini-plug inputs, makes the TC-100 an accurate, easy-to-use instrument for all your thermocouple calibration needs.

General Specifications:
(1 year at 23°C ±5°C; % of reading unless otherwise noted)
Output Voltage
  Range: -10 to +75.000 mV
Resolution: 1 µV
Accuracy: ±0.007 % of reading, ±10 µV
Output Impedance: &Mac178;1 Ohm
Input Voltage:
  Range: -10 to +75.000 mV
Resolution: 1 µV
Accuracy: ±0.007 % of reading, ±10 µV
Output Impedance: Input Impedance > 1 MegOhm
Thermocouple Source/Measure:
  Types: J, K, T, E, R, S, B, L, U, C
Range: mV
Resolution
   Source: ±0.1°C or °F
   Measure ±0.01°C or °F
Accuracy:
J ±0.5 °C; -210 °C to -100 °C
±0.3 °C; -100 °C to +1,200 °C
K ±0.6 °C; -200 °C to -100 °C
±0.35 °C; -100 °C to +1,000 °C
±0.5 °C; +1,000 °C to +1,372 °C
T ±0.7 °C; -200 °C to -150 °C
±0.3 °C; -150 °C to +400 °C
E ±0.5 °C; -200 °C to -100 °C
±0.3 °C; -100 °C to +1,000 °C
R ±1.8 °C; 0 °C to 250 °C
±1.0 °C; 250 °C to +1,767 °C
S ±1.8 °C; 0 °C to 250 °C
±1.0 °C; 250 °C to +1,767 °C
B ±1.7 °C; 600 °C to 1,000 °C
±1.2 °C; 1,000 °C to 1,820 °C
L ±0.5 °C; -200 °C to -100 °C
±0.4 °C; -100 °C to +900 °C
U ±0.7 °C; -200 °C to 0 °C
±0.3 °C; 0 °C to +600 °C
C ±0.4 °C; 0 to °C 1,000 °C
±0.7 °C; 1,000 °C to +1,800 °C
±1.2 °C; +1,800 °C to +2,316 °C
CJC Temperature Offset: ±0.05 °C/°C outside of 23±5 °C
Warm-up Time: 1 minute to specification
Environmental:
  Operating: -10 °C to +55 °C
Storage: -20 °C to +70 °C
Power Requirements:
Battery: 9 Volt alkaline; 006P/ IEC 6F22/ NEDA1604
Optional NiCad
Optional AC adapter/charger
Mechanical Dimensions: 5.7" x 3.15" x 1.43"
(14.47 cm x 8.00 cm x 3.63 cm)
Weight: 12 ounces (340 grams)
Notes:
 1. Temperature standard ITS-90.

 

 

Calibrators | Pressure | Temperature | Loop | Voltage & Frequency | Multi-Function | Bench

Other | Process Displays | Power Supplies

 

To Place an Order, or to Receive any Additional Information, Please e-mail:  process@bcgroupintl.com, call (314) 638-3800 or fax (314) 638-3200