PTC-8001
Martel
Home
High Accuracy Temperature Calibrator "Smart"
Transmitter Compatible |
|
|
|
|
 |
|
COMMON FEATURES
|
|
High
accuracy ±0.4°C J T/C and ±0.3°C 4W PT385 RTD all errors
included |
|
10 T/C
types and 8 RTD types including PT 385 (1000 ohm) and cu (10 ohm) |
|
Direct
keyboard entry or scroll control of output |
|
Nema 4
rated case, rugged design |
|
RTD
simulation works with all pulsed (smart) transmitters |
|
YSI 400
series thermistor range for medical applications |
|
Standard mini-plug or banana jack connections for T/C |
|
Store up to nine setpoints for each output function |
|
RS232 Interface |
|
The Martel PTC-8001
Precision Temperature Calibrator is an ideal device to handle all of your
temperature calibration needs in a truly rugged, low cost package. It combines
virtually all widely used thermocouples and RTDs in one deviceyoull
never have to worry about finding a temperature device you cant
calibrate.
This is especially true when calibrating smart
or pulsed RTD transmitters where many other calibrators fail to work or
operate at reduced accuracy specs. The PTC-8001 handles those applications with
ease.
Each Unit Includes: Test leads, carrying case, 4 AA alkaline
batteries, NIST Certificate, and instruction manual.
Note: For T/C
please order T/C cable kit. |
|
General Specifications: |
|
T/C Ranges: |
J, K, T, E, R, S, N, B, L, U including 10 to 70 mV
range |
|
RTD Ranges: |
Pt 385 (100, 200, 500, 1000 ohms) Pt 392, JIS, Ni 120, CU10,
YSI 400 |
|
Ohms Ranges: |
0 to 400.00 and 400.0 to 4000.0 |
|
Accuracy: |
T/C Type J: ±.3°C ±10
 V (±0.6°C
total error) RTD PT100: ±.1°C ±.075
(±0.3°C total error) Ohms (400):
±0.1 Ohms (4000): ±1.0
 |
|
RTD IEX-range: |
0.01 mA to 3 mA |
|
RTD Frequency Response: |
10 mS works with all pulsed XMTRs |
|
Oper. Temp: |
-10°C to 50°C |
|
Storage Temp: |
-20°C to 60°C |
|
Power Supply: |
4 AA Alkaline Cells, 30 hours typical operating
time |
|
Size: |
7.5" x 4" x 1.5" |
|
|
Range: |
Min |
Max |
Accuracy |
|
|
Ohms Read 4w
(low): |
0.00 |
400.00 |
0.1 ohm |
|
|
Ohms Read 4w
(high): |
400.0 1500.0 |
1500.0 4000.0 |
0.5 ohm 1.0 ohm |
|
|
Ohms
Ranges: |
|
|
|
|
|
Range: |
Min |
Max |
Excitation Current |
Accuracy |
|
Ohms Source
(low): |
5.00 5.00 |
400.00 400.00 |
0.1 to 0.5 mA 0.5 to 3 mA |
0.15 ohm 0.1 ohm |
|
Ohms Source
(high): |
400.0 1500.0 |
1500.0 4000.0 |
0.05 to 0.8 mA 0.05 to 0.4 mA |
0.5 ohm 1.0 ohm |
|
|
RTD Read and Source
|
|
Range: |
Min |
Max |
Accuracy |
|
Ni120 (672) |
-80.0 |
260.0 |
0.2°C |
|
Pt100 (385) |
-200.0 |
800.0 |
0.33°C |
|
Pt100(3926) |
-200.0 |
630.0 |
0.3°C |
|
Pt100(3916) |
-200.0 |
630.0 |
0.3°C |
|
Pt200(385) |
-200.0 100.0 |
630.0 |
0.8°C |
|
Pt500(385) |
-200.0 500.0 |
500.0 630.0 |
0.3°C 0.4°C |
|
Pt1000(385) |
-200.0 100.0 |
100.0 630.0 |
0.2°C 0.3°C |
|
Cu10 |
-100.0 |
250.0 |
2.2°C |
|
YSI400 |
15.00 |
50.00 |
0.05°C |
|
|
Thermocouple Read and Source (All errors included)
|
|
Range: |
Min |
Max |
Accuracy |
|
J Thermocouple |
-200.0 0.0 |
0.0 1200.0 |
0.6°C 0.4°C |
|
K Thermocouple |
-200.0 0.0 |
0.0 1370.0 |
0.8°C 0.5°C |
|
T Thermocouple |
-200.0 0.0 |
0.0 400.0 |
0.8°C 0.5°C |
|
E Thermocouple |
-200.0 0.0 |
0.0 950.0 |
0.5°C 0.4°C |
|
R Thermocouple |
-20 0 50.0 |
0.0 500 1750 |
2.4°C 1.7°C 1.3°C |
|
|
Range: |
Min. |
Max. |
Accuracy
|
|
S Thermocouple |
-20 0 500 |
-20 0 1750 |
2.4°C 1.7°C 1.4°C |
|
B Thermocouple |
600 8000 1000 |
800 1000 1800 |
2.1°C 1.7°C 1.3°C |
|
L Thermocouple |
-200.0 0.00 |
0.0 900.0 |
0.45°C 0.4°C |
|
U Thermocouple |
-200.0 0.00 |
0.0 400.0 |
0.7°C 0.45°C |
|
N Thermocouple |
-200.0 0.00 |
0.0 1300.0 |
1.1°C 0.6°C |
|
mV Read/Source |
-10.00 |
75.00 |
0.015% ± 2 |
|
|
T/C Cable Kit |
|
|
|
KIT INCLUDES: |
 |
|
Type J
T/C cable |
|
Type T
T/C cable |
|
Type K
T/C cable |
|
Type E
T/C cable |
|
Ideal for use with
MC-1000, PTC-8001 and T120 calibrators. Martel part number 80029.
All
cables are terminated with a standard mini-connector on one end and
bare leads
on the other. |
|
|