MC-1000
Martel
Home
High Accuracy Multi-Function
Calibrator |
|
|
|
COMMON FEATURES |
 |
|
Excellent accuracy ±0.4°C T/C ±0.3°C RTD and
0.015% electrical |
|
Choose
either read or source modes |
|
Operates
with external pressure modules |
|
10 T/C
types; 8 RTD types plus YSI 400 series thermistor (medical
applications) |
|
Full
function 4 to 20 mA read, source, simulate and read/power modes with 0.015%
reading accuracy |
|
Source 0
to 20 volts or read 0 to 20 volts with 0.015% of reading |
|
Measure 0 to 250 VDC or VAC |
|
Store up to nine setpoints for each |
|
RTD function works with all pulsed XMTRs |
|
Direct keyboard entry allows for quick output changes |
|
RS232 Interface |
A
full multi-function calibrator at a very competitive price
with excellent accuracy best describes the MC-1000. The
MC-1000 combines the accuracy and functions of our next
generation PTC-8000 temperature calibrator with V/mA and
frequency functions added. The performance level is twice that of similarly priced
products and will calibrate virtually any process device.
The MC-1000
also combines several DMM functions like voltage measurement to 250 V AC or DC
and a continuity test function. This combination of a high accuracy calibrator
along with common DMM functions makes the MC-1000 the only instrument you may
ever need to maintain and calibrate your instrumentation.
Each unit
includes: Test leads, carrying case, 4 AA alkaline batteries, NIST Certificate,
and instruction manual. |
|
General Specifications: |
|
Pressure Ranges: |
Various ranges in 29 different pressure modules see
Pressure Module Ranges: Specifications |
|
T/C Ranges: |
J,K,T,E,R,S,N,B,L,U including -10 to70 mV |
|
RTD Ranges: |
Pt 385 (100, 200, 500, 1000 ohms) Pt 392, JIS, Ni120, CU10,
YSI400 |
|
Ohms Ranges: |
0 to 400.00 and 400.0 to 4000.0 |
|
mA Range: |
0 to 24.000 mA |
|
Voltage: |
(Low) Read 0-20.000 VDC Source 0-10.000 VDC (High) Read only
0-250 VDC or VAC |
|
Frequency: |
0-1000.0 CPM 0-1000.0 Hz 0-10.00 KHz |
|
Continuity: |
0 to 400.0 ohms, audible tone when under 10 ohms |
|
Accuracy: |
T/C Type J: ±.3°C ±10 m V
(±0.4°C total error) RTD PT100: ±.1°C
±.075 (±0.3°C total error) Ohms (400): ±0.1
Ohms (4000): ±1.0
 |
|
mA Compliance: |
Up to 1000
load |
|
Voltage Drive Capability: |
1 mA |
|
RTD IEX-range: |
0.01 mA to 3 mA |
|
RTD Frequency Response: |
10 mS works with all pulsed XMTRs |
|
Oper. Temp: |
-10°C to 50°C |
|
Storage Temp: |
-20°C to 70°C |
|
Loop Power: |
4 AA alkaline batteries, 25 hours typical operating
time |
|
Size: |
7.5" x 4" x 1.5" |
|
|
|